1024 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:NRZ:DSIZe
(see page 1610)
Command Syntax
:SBUS<n>:NRZ:DSIZe <#words>
<#words> ::= from 1-255, in NR1 format
The :SBUS<n>:NRZ:DSIZe command ...
When the display format is WORD (see :SBUS<n>:NRZ:DISPlay), the
:SBUS<n>:NRZ:DSIZe command specifies the number of words in the data field of
your NRZ protocol definition.
AUTO is available as a selection only when the trailer field size is 0 (see
:SBUS<n>:NRZ:TSIZe).
Query Syntax
:SBUS<n>:NRZ:DSIZe?
The :SBUS<n>:NRZ:DSIZe? query returns the number of data field words setting.
Return Format
<#words><NL>
<#words> ::= from 1-255, in NR1 format
In AUTO mode, the query returns 0.
In BIT format the only legal value is 0 (for AUTO).
See Also • ":SBUS<n>:NRZ:DISPlay" on page 1023
• ":SBUS<n>:NRZ:FSIZe" on page 1025
• ":SBUS<n>:NRZ:HSIZe" on page 1026
• ":SBUS<n>:NRZ:STARt" on page 1031
• ":SBUS<n>:NRZ:TSIZe" on page 1035
• ":SBUS<n>:NRZ:WSIZe" on page 1036