EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1462 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:MODulation:STATe
(see page 1610)
Command Syntax
:WGEN<w>:MODulation:STATe <setting>
<w> ::= 1 in NR1 format
<setting> ::= {{OFF | 0} | {ON | 1}}
The :WGEN<w>:MODulation:STATe command enables or disables modulated
waveform generator output.
You can enable modulation for all waveform generator function types except
pulse, DC, and noise.
Query Syntax
:WGEN<w>:MODulation:STATe?
The :WGEN<w>:MODulation:STATe? query returns whether the modulated
waveform generator output is enabled of disabled.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also ":WGEN<w>:MODulation:AM:DEPTh" on page 1453
":WGEN<w>:MODulation:AM:FREQuency" on page 1454
":WGEN<w>:MODulation:FM:DEViation" on page 1455
":WGEN<w>:MODulation:FM:FREQuency" on page 1456
":WGEN<w>:MODulation:FSKey:FREQuency" on page 1457
":WGEN<w>:MODulation:FSKey:RATE" on page 1458
":WGEN<w>:MODulation:FUNCtion" on page 1459
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1460
":WGEN<w>:MODulation:TYPE" on page 1463

Table of Contents

Related product manuals