:MTESt Commands 26
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 687
:MTESt:RMODe:FACTion:SAVE
(see page 1610)
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures on or
off.
See Chapter 30, “:SAVE Commands,” starting on page 821 for more information
on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure save
setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also • "Introduction to :MTESt Commands" on page 667
• ":MTESt:RMODe:FACTion:MEASure" on page 685
• ":MTESt:RMODe:FACTion:PRINt" on page 686
• ":MTESt:RMODe:FACTion:STOP" on page 688
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets :MTESt:RMODe:FACTion:PRINt
OFF.