362 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
11 :CHANnel<n> Commands
:CHANnel<n>:PROTection
(see page 1610)
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n> ::= 1 to (# analog channels) in NR1 format| 4}
When the analog channel input impedance is set to 50Ω, the input channels are
protected against overvoltage. When an overvoltage condition is sensed, the input
impedance for the channel is automatically changed to 1 M
Ω.
The :CHANnel<n>:PROTection[:CLEar] command is used to clear (reset) the
overload protection. It allows the channel to be used again in 50
Ω mode after the
signal that caused the overload has been removed from the channel input.
Reset the analog channel input impedance to 50
Ω (see
":CHANnel<n>:IMPedance" on page 350) after clearing the overvoltage
protection.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query returns the state of the input protection for
CHANnel<n>. If a channel input has experienced an overload, TRIP (tripped) will
be returned; otherwise NORM (normal) is returned.
Return Format
{NORM | TRIP}<NL>
See Also • "Introduction to :CHANnel<n> Commands" on page 345
• ":CHANnel<n>:COUPling" on page 348
• ":CHANnel<n>:IMPedance" on page 350
• ":CHANnel<n>:PROBe" on page 354

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals