862 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:A429:BAUDrate
(see page 1610)
Command Syntax
:SBUS<n>:A429:BAUDrate <baudrate>
<baudrate> ::= integer from 10000 to 1000000
When a user-defined baud rate is selected (with the ":SBUS<n>:A429:SPEed
USER" command), the :SBUS<n>:A429:BAUDrate command specifies the
user-defined baud rate. The baud rate can be set in 100 b/s increments between
10000 and 100000 and in 1000 b/s ingrements between 100000 and 1000000.
Query Syntax
:SBUS<n>:A429:BAUDrate?
The :SBUS<n>:A429:BAUDrate? query returns the user-defined baud rate setting.
Return Format
<baudrate><NL>
See Also ":SBUS<n>:A429:SPEed" on page 869