:SEARch Commands 32
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1195
:SEARch:SERial:CAN:SYMBolic:VALue
(see page 1610)
Command Syntax
:SEARch:SERial:CAN:SYMBolic:VALue <data>
<data> ::= value in NR3 format
The :SEARch:SERial:CAN:SYMBolic:VALue command specifies the signal value to
search for when CAN symbolic data has been loaded (recalled) into the
oscilloscope and the CAN serial search mode is set to MSIGnal.
Query Syntax
:SEARch:SERial:CAN:SYMBolic:VALue?
The :SEARch:SERial:CAN:SYMBolic:VALue? query returns the specified signal
value.
Return Format
<data><NL>
<data> ::= value in NR3 format
See Also ":RECall:DBC[:STARt]" on page 813
":SEARch:SERial:CAN:MODE" on page 1187
":SEARch:SERial:CAN:SYMBolic:MESSage" on page 1193
":SEARch:SERial:CAN:SYMBolic:SIGNal" on page 1194
NOTE
Encoded signal values are not supported in the remote interface (even though they can be
used in the front panel graphical interface).