:POWer Commands 28
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 771
:POWer:PSRR:WGEN:LOAD
(see page 1610)
Command Syntax
:POWer:PSRR:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :POWer:PSRR:WGEN:LOAD command sets the waveform generator expected
output load impedance.
The output impedance of the Gen Out signal is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load. If the actual load impedance is
different than the selected value, the displayed amplitude and offset levels will be
incorrect.
Query Syntax
:POWer:PSRR:WGEN:LOAD?
The :POWer:PSRR:WGEN:LOAD? query returns the waveform generator expected
output load impedance setting.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also • ":POWer:PSRR" on page 760
• ":POWer:PSRR:APPLy" on page 761
• ":POWer:PSRR:DATA" on page 762
• ":POWer:PSRR:FREQuency:MAXimum" on page 763
• ":POWer:PSRR:FREQuency:MINimum" on page 764
• ":POWer:PSRR:FREQuency:MODE" on page 765
• ":POWer:PSRR:PPDecade" on page 767
• ":POWer:PSRR:SOURce:INPut" on page 768
• ":POWer:PSRR:SOURce:OUTPut" on page 769
• ":POWer:PSRR:WGEN:VOLTage" on page 772
• ":POWer:PSRR:WGEN:VOLTage:PROFile" on page 773