792 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:SIGNals:VSTeady:TRANsient
(see page 1610)
Command Syntax
:POWer:SIGNals:VSTeady:TRANsient <value>[suffix]
<value> ::= Expected steady state output Voltage value in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VSTeady:TRANsient command specifies the expected steady
state output DC voltage of the power supply for transient response analysis.
This value is used along with the overshoot percentage to specify the settling band
for the transient response and to adjust the vertical scale of the oscilloscope.
Query Syntax
:POWer:SIGNals:VSTeady:TRANsient?
The :POWer:SIGNals:VSTeady:TRANsient query returns the expected steady state
voltage setting.
Return Format
<value><NL>
<value> ::= Expected steady state output Voltage value in NR3 format
See Also • ":POWer:TRANsient:APPLy" on page 803
• ":POWer:SIGNals:AUTosetup" on page 776
• ":POWer:SIGNals:DURation:TRANsient" on page 784
• ":POWer:SIGNals:IEXPected" on page 785
• ":POWer:SIGNals:OVERshoot" on page 786
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 793
• ":POWer:SIGNals:SOURce:VOLTage<i>" on page 794

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals