1422 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
36 :WAVeform Commands
:WAVeform:TYPE
(see page 1610)
Query Syntax
:WAVeform:TYPE?
The :WAVeform:TYPE? query returns the acquisition mode associated with the
currently selected waveform. The acquisition mode is set by the :ACQuire:TYPE
command.
Return Format
<mode><NL>
<mode> ::= {NORM | PEAK | AVER | HRES}
See Also • "Introduction to :WAVeform Commands" on page 1396
• ":ACQuire:TYPE" on page 318
• ":WAVeform:DATA" on page 1403
• ":WAVeform:PREamble" on page 1410
• ":WAVeform:SOURce" on page 1417
If the :WAVeform:SOURce is POD1, POD2, or SBUS1, SBUS2, the type is always NORM.