EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #374 background imageLoading...
Page #374 background image
374 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
12 :COMPliance Commands
:COMPliance:USB:SOURce:DPLus
(see page 1610)
Command Syntax
:COMPliance:USB:SOURce:DPLus <source>
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :COMPliance:USB:SOURce:DPLus command specifies the analog input
channel that is probing the D+ signal.
On 4-channel oscilloscopes, you are forced to use different channel pairs for the
D+ and D- signals. This provides the maximum sample rate. (Channels 1 and 2 are
one pair and channels 3 and 4 are the other pair.)
Query Syntax
:COMPliance:USB:SOURce:DPLus?
The :COMPliance:USB:SOURce:DPLus? query returns the specified analog input
channel.
Return Format
<source><NL>
<source> ::= {CHAN<n>}
See Also • ":COMPliance:USB:SOURce:DMINus" on page 373
• ":COMPliance:USB:SOURce:ADJacent" on page 371

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish