:CALibrate Commands 10
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 335
:CALibrate:OUTPut
(see page 1610)
Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | WAVEgen | WGEN1 | WGEN2 | NFC | TSOurce}
Note: WAVE and WGEN1 are equivalent.
Note: WGEN2 only available on models with 2 WaveGen outputs.
The CALibrate:OUTPut command sets the signal that is available on the rear panel
TRIG OUT BNC:
• TRIGgers — pulse when a trigger event occurs. The output level is 0-5 V into an
open circuit, and 0-2.5 V into 50
Ω.
• MASK — signal from mask test indicating a failure.
• WAVEgen, WGEN1, WGEN2 — waveform generator sync output signal. This
signal depends on the :WGEN<w>:FUNCtion setting:
• NFC — This option is available in the Near Field Communication (NFC) trigger
mode when the ATRigger (Arm & Trigger) trigger event is selected (see
":TRIGger:NFC:TEVent" on page 1345). The ATRigger trigger event lets you arm
the oscilloscope on one event and then trigger on a second event or after a
specified timeout period if the second event does not occur.
When :CALibrate:OUTPut is NFC and the specified event arms, the TRIG
OUTBNC goes high. The oscilloscope waits until a second event is found or
until the specified timeout period expires and then triggers.
• For NFC-A, the second event is SDD_REQ.
• For NFC-B, the second event is ATTRIB.
• For NFC-F, the second event is ATR_REQ.
When the oscilloscope triggers, the TRIG OUTBNC line goes low.
Waveform
Type
Sync Signal Characteristics
SINusoid,
SQUare,
RAMP, PULSe,
SINC,
EXPRise,
EXPFall,
GAUSsian
The Sync signal is a TTL positive pulse that occurs when the waveform rises above
zero volts (or the DC offset value).
DC, NOISe,
CARDiac
N/A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals