336 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
10 :CALibrate Commands
TSOurce — The raw trigger signal from the oscilloscope's trigger circuit is
output to Trig Out. It produces a rising edge whenever the input source would
cause a trigger, even though that might occur multiple times within the time of
a single acquisition. The trigger source can be a front panel analog input
channel or an external trigger input. The output level is 0-5 V into an open
circuit, and 0-2.5 V into 50
Ω. This option is not available in all trigger modes.
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the current source of the TRIG OUT BNC
signal.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | WAVE | WGEN2 | NFC | TSO}
See Also "Introduction to :CALibrate Commands" on page 332
":WGEN<w>:FUNCtion" on page 1446