1094 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:UART:BAUDrate
(see page 1610)
Command Syntax
:SBUS<n>:UART:BAUDrate <baudrate>
<baudrate> ::= integer from 100 to 8000000, 10000000, or 12000000
The :SBUS<n>:UART:BAUDrate command selects the bit rate (in bps) for the serial
decoder and/or trigger when in UART mode. The baud rate can be set in the range
from from 100 b/s to 8 Mb/s or to the specific values of 10 Mb/s or 12 Mb/s.
If the baud rate you select does not match the system baud rate, false triggers
may occur.
Query Syntax
:SBUS<n>:UART:BAUDrate?
The :SBUS<n>:UART:BAUDrate? query returns the current UART baud rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 100 to 8000000, 10000000, or 12000000
See Also "Introduction to :TRIGger Commands" on page 1295