1230 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
32 :SEARch Commands
:SEARch:SERial:SENT:MODE
(see page 1610)
Command Syntax
:SEARch:SERial:SENT:MODE <mode>
<mode> ::= {FCData | SCMid | SCData | CRCerror | PPERror}
When SENT serial decode is turned on and displayed in the Lister, the
:SEARch:SERial:SENT:MODE command specifies what to search for in the
decoded data:
FCData — finds Fast Channel data nibbles that match the values entered using
additional softkeys.
SCMid — finds Slow Channel Message IDs that match the value entered using
additional softkeys.
SCData — finds Slow Channel Message IDs and Data that match the values
entered using additional softkeys.
CRCerror — finds any CRC error, Fast or Slow.
PPERror — finds where a nibble is either too wide or too narrow (for example,
data nibble < 12 (11.5) or > 27 (27.5) ticks wide). Sync, S&C, data, or checksum
pulse periods are checked.
Query Syntax
:SEARch:SERial:SENT:MODE?
The :SEARch:SERial:SENT:MODE? query returns the search mode setting.
Return Format
<mode><NL>
<mode> ::= {FCD | SCM | SCD | CRC | PPER}
See Also ":SEARch:SERial:SENT:FAST:DATA" on page 1229
":SEARch:SERial:SENT:SLOW:DATA" on page 1231
":SEARch:SERial:SENT:SLOW:ID" on page 1232