:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1133
:SBUS<n>:USB:TRIGger:PID:TOKen
(see page 1610)
Command Syntax
:SBUS<n>:USB:TRIGger:PID:TOKen <pid>
<pid> ::= {OUT | IN | SETup | SOF}
The :SBUS<n>:USB:TRIGger:PID:TOKen command specifies the USB token PID to
trigger on for the selected bus.
The specified PID does not include the PID check value, which is specified using
the :SBUS<n>:USB:TRIGger:PIDCheck command.
The TOKen USB trigger mode can be selected using the :SBUS<n>:USB:TRIGger
command.
Query Syntax
:SBUS<n>:USB:TRIGger:PID:TOKen?
The :SBUS<n>:USB:TRIGger:PID:TOKen? query returns the specified token PID.
Return Format
<pid><NL>
<pid> ::= {OUT | IN | SETup | SOF}
See Also • ":SBUS<n>:USB:TRIGger:PCHeck" on page 1129
• ":SBUS<n>:USB:TRIGger" on page 1120