918 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth
(see page 1610)
Command Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer between 0 and 12, in NR1 format.
For the trigger types that let you trigger on data, the
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth command specifies the length of
the data to trigger on, from 0 to 12 bytes, limited by the data length code (DLC)
setting of the :SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC command.
The :SBUS<n>:CXPI:TRIGger:PATTern:DATA command lets you specify the data
value to trigger on.
Query Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth? query returns the data
length setting.
Return Format
<length><NL>
<length> ::= integer between 0 and 12, in NR1 format.
See Also • ":SBUS<n>:CXPI:TRIGger" on page 913
• ":SBUS<n>:CXPI:TRIGger:IDFilter" on page 915
• ":SBUS<n>:CXPI:TRIGger:PTYPe" on page 916
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA" on page 917
• ":SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt" on page 919
• ":SBUS<n>:CXPI:TRIGger:PATTern:ID" on page 920
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:CT" on page 921
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC" on page 922
• ":SBUS<n>:CXPI:TRIGger:PATTern:INFO:NM" on page 923