Obsolete and Discontinued Commands 39
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1507
:EXTernal:PMODe
(see page 1610)
Command Syntax
:EXTernal:PMODe <pmode value>
<pmode value> ::= {AUTo | MANual}
The probe sense mode is controlled internally and cannot be set. If a probe with
sense is connected to the specified channel, auto sensing is enabled; otherwise,
the mode is manual.
If the pmode sent matches the oscilloscope's setting, the command will be
accepted. Otherwise, a setting conflict error is generated.
Query Syntax
:EXTernal:PMODe?
The :EXTernal:PMODe? query returns AUT if an autosense probe is attached and
MAN otherwise.
Return Format
<pmode value><NL>
<pmode value> ::= {AUT | MAN}
NOTE
The :EXTernal:PMODe command is an obsolete command provided for compatibility to
previous oscilloscopes.