382 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
13 :COUNter Commands
:COUNter:MODE
(see page 1610)
Command Syntax
:COUNter:MODE <mode>
<mode> ::= {FREQuency | PERiod}
The :COUNter:MODE command sets the counter mode:
FREQuency — the cycles per second (Hz) of the signal.
PERiod — the time periods of the signal's cycles.
Query Syntax
:COUNter:MODE
The :COUNter:MODE? query returns the counter mode setting.
Return Format
<mode><NL>
<mode> ::= {FREQ | PER}
See Also ":COUNter:CURRent" on page 380
":COUNter:ENABle" on page 381
":COUNter:NDIGits" on page 383
":COUNter:SOURce" on page 384