:MEASure Power Commands 25
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 655
:MEASure:ONTime
(see page 1610)
Command Syntax
:MEASure:ONTime [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:ONTime command installs an "on time" measurement on screen.
Turn on time measures the difference of time between when the input AC Voltage
first rises to 10% of its maximum amplitude to the time when the output DC
Voltage rises to 90% of its maximum amplitude.
The <source1> parameter is the AC Voltage and the <source2> parameter is the
DC Voltage. These sources can also be specified by the :MEASure:SOURce
command.
Query Syntax
:MEASure:ONTime? [<source1>][,<source2>]
The :MEASure:ONTime query returns the measured turn off time.
Return Format
<return_value><NL>
<return_value> ::= the time in seconds in NR3 format
See Also ":MEASure:SOURce" on page 615
":POWer:ONOFf:TEST" on page 757
":POWer:ONOFf:APPLy" on page 754