:TRIGger Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1349
:TRIGger:NFC:TIMeout:TIME
(see page 1610)
Command Syntax
:TRIGger:NFC:TIMeout:TIME <time>
<time> ::= seconds in NR3 format
The :TRIGger:NFC:TIMeout:TIME command specifies the timeout period. With the
ATRigger (Arm & Trigger) trigger event, if the second event does not occur within
the timeout period, the oscilloscope triggers when the timeout period expires.
Query Syntax
:TRIGger:NFC:TIMeout:TIME?
The :TRIGger:NFC:TIMeout:TIME? query returns the specified timeout period.
Return Format
<time><NL>
<time> ::= seconds in NR3 format
See Also • ":TRIGger:NFC:AEVent" on page 1340
• ":TRIGger:NFC:ATTime" on page 1341
• ":TRIGger:NFC:RPOLarity" on page 1342
• ":TRIGger:NFC:SOURce" on page 1343
• ":TRIGger:NFC:STANdard" on page 1344
• ":TRIGger:NFC:TEVent" on page 1345
• ":TRIGger:NFC:TIMeout" on page 1347
• ":TRIGger:NFC:TIMeout:ENABle" on page 1348

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals