1130 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:USB:TRIGger:PID:DATA
(see page 1610)
Command Syntax
:SBUS<n>:USB:TRIGger:PID:DATA <pid>
<pid> ::= {DATA0 | DATA1 | DATA2 | MDATa}
The :SBUS<n>:USB:TRIGger:PID:DATA command specifies the USB data PID to
trigger on for the selected bus.
The specified PID does not include the PID check value, which is specified using
the :SBUS<n>:USB:TRIGger:PIDCheck command.
The DATA USB trigger mode can be selected using the :SBUS<n>:USB:TRIGger
command.
Query Syntax
:SBUS<n>:USB:TRIGger:PID:DATA?
The :SBUS<n>:USB:TRIGger:PID:DATA? query returns the specified data PID.
Return Format
<pid><NL>
<pid> ::= {DATA0 | DATA1 | DATA2 | MDAT}
See Also ":SBUS<n>:USB:TRIGger:PCHeck" on page 1129
":SBUS<n>:USB:TRIGger" on page 1120