Obsolete and Discontinued Commands 39
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1503
:CHANnel<n>:PMODe
(see page 1610)
Command Syntax
:CHANnel<n>:PMODe <pmode value>
<pmode value> ::= {AUTo | MANual}
<n> ::= 1 to (# analog channels) in NR1 format
The probe sense mode is controlled internally and cannot be set. If a probe with
sense is connected to the specified channel, auto sensing is enabled; otherwise,
the mode is manual.
If the PMODe sent matches the oscilloscope's setting, the command will be
accepted. Otherwise, a setting conflict error is generated.
Query Syntax
:CHANnel<n>:PMODe?
The :CHANnel<n>:PMODe? query returns AUT if an autosense probe is attached
and MAN otherwise.
Return Format
<pmode value><NL>
<pmode value> ::= {AUT | MAN}
NOTE
The :CHANnel<n>:PMODe command is an obsolete command provided for compatibility to
previous oscilloscopes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals