:TRIGger Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1343
:TRIGger:NFC:SOURce
(see page 1610)
Command Syntax
:TRIGger:NFC:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:NFC:SOURce command selects the input waveform source for the
NFC trigger. You can choose an analog input channel.
Query Syntax
:TRIGger:NFC:SOURce?
The :TRIGger:NFC:SOURce? query returns the input waveform source setting.
Return Format
<source><NL>
<source> ::= CHAN<n>
See Also • ":TRIGger:NFC:AEVent" on page 1340
• ":TRIGger:NFC:ATTime" on page 1341
• ":TRIGger:NFC:RPOLarity" on page 1342
• ":TRIGger:NFC:STANdard" on page 1344
• ":TRIGger:NFC:TEVent" on page 1345
• ":TRIGger:NFC:TIMeout" on page 1347
• ":TRIGger:NFC:TIMeout:ENABle" on page 1348
• ":TRIGger:NFC:TIMeout:TIME" on page 1349