EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #870 background imageLoading...
Page #870 background image
870 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:A429:TRIGger:LABel
(see page 1610)
Command Syntax
:SBUS<n>:A429:TRIGger:LABel <value>
<value> ::= 8-bit integer in decimal, <hex>, <octal>, or <string>
from 0-255 or "0xXX" (don't care)
<hex> ::= #Hnn where n ::= {0,..,9 | A,..,F}
<octal> ::= #Qnnn where n ::= {0,..,7}
<string> ::= "0xnn" where n::= {0,..,9 | A,..,F}
The :SBUS<n>:A429:TRIGger:LABel command defines the ARINC 429 label value
when labels are used in the selected trigger type.
To set the label value to don't cares (0xXX), set the value to -1.
Query Syntax
:SBUS<n>:A429:TRIGger:LABel?
The :SBUS<n>:A429:TRIGger:LABel? query returns the current label value in
decimal format.
Return Format
<value><NL> in decimal format
Errors • "-241, Hardware missing" on page 1551
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:A429:TRIGger:TYPE" on page 875

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals