1372 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:TRANsition:QUALifier
(see page 1610)
Command Syntax
:TRIGger:TRANsition:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan}
The :TRIGger:TRANsition:QUALifier command specifies whether you are looking for
rise/fall times greater than or less than a certain time value. The time value is set
using the :TRIGger:TRANsition:TIME command.
Query Syntax
:TRIGger:TRANsition:QUALifier?
The :TRIGger:TRANsition:QUALifier? query returns the current rise/fall time trigger
qualifier setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:TRANsition:TIME" on page 1375
• ":TRIGger:MODE" on page 1308