:TIMebase Commands 34
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1285
:TIMebase:RANGe
(see page 1610)
Command Syntax
:TIMebase:RANGe <range_value>
<range_value> ::= time for 10 div in seconds in NR3 format
The :TIMebase:RANGe command sets the full-scale horizontal time in seconds for
the main window. The range is 10 times the current time-per-division setting.
Query Syntax
:TIMebase:RANGe?
The :TIMebase:RANGe query returns the current full-scale range value for the
main window.
Return Format
<range_value><NL>
<range_value> ::= time for 10 div in seconds in NR3 format
See Also "Introduction to :TIMebase Commands" on page 1282
":TIMebase:MODE" on page 1283
":TIMebase:SCALe" on page 1289
":TIMebase:WINDow:RANGe" on page 1292
Example Code
' TIME_RANGE - Sets the full scale horizontal time in seconds. The
' range value is 10 times the time per division.
myScope.WriteString ":TIM:RANG 2e-3" ' Set the time range to 0.002
seconds.
See complete example programs at: Chapter 44, “Programming Examples,”
starting on page 1619

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals