1390 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:ZONE<n>:PLACement
(see page 1610)
Command Syntax
:TRIGger:ZONE<n>:PLACement <width>, <height>, <x_center>, <y_center>
<width> ::= width of zone in seconds
<height> ::= height of zone in volts
<x_center> ::= center of zone in seconds
<y_center> ::= center of zone in volts
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:PLACement command sets the size and location of Zone 1
or Zone 2.
No error is returned if the zone is placed off-screen, or if the zones overlap such
that Zone 2 becomes invalid. The :TRIGger:ZONE<n>:VALidity? query is used to
retrieve this information.
Query Syntax
:TRIGger:ZONE<n>:PLACement?
The :TRIGger:ZONE<n>:PLACement? query returns the size and location of Zone 1
or Zone 2.
Return Format
<opt><NL>
<opt> ::= <width>, <height>, <x_center>, <y_center>
See Also ":TRIGger:ZONE<n>:STATe" on page 1392
":TRIGger:ZONE<n>:MODE" on page 1389
":TRIGger:ZONE<n>:VALidity" on page 1391