EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Obsolete and Discontinued Commands 39
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1523
:MEASure:TMIN
(see page 1610)
Command Syntax
:MEASure:TMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMIN command installs a screen measurement and starts an
X-at-Min-Y measurement on the selected waveform. If the optional source is
specified, the current source is modified.
Query Syntax
:MEASure:TMIN? [<source>]
The :MEASure:TMIN? query returns the horizontal axis value at which the minimum
vertical value occurs on the current source. If the optional source is specified, the
current source is modified.
Return Format
<value><NL>
<value> ::= time at minimum in NR3 format
See Also "Introduction to :MEASure Commands" on page 568
":MEASure:TMAX" on page 1522
":MEASure:XMAX" on page 638
":MEASure:XMIN" on page 639
NOTE
The :MEASure:TMIN command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMIN command (see page 639) instead.

Table of Contents

Related product manuals