:SAVE Commands 30
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 841
:SAVE:RESults:FORMat:MEASurement
(see page 1610)
Command Syntax
:SAVE:RESults:FORMat:MEASurement {{0 | OFF} | {1 | ON}}
The :SAVE:RESults:FORMat:MEASurement command specifies whether
measurement results will be included when analysis results are saved.
Analysis results are saved using the :SAVE:RESults:[STARt] command.
Other :SAVE:RESults:FORMat commands specify whether other types of analysis
results are also saved.
When multiple types of analysis results are saved, they are all saved to the same
file and separated by a blank line.
Query Syntax
:SAVE:RESults:FORMat:MEASurement?
The :SAVE:RESults:FORMat:MEASurement? query returns whether measurement
results will be included when analysis results are saved.
Return Format
<off_on><NL>
{0 | 1}
See Also • ":SAVE:RESults:[STARt]" on page 838
• ":SAVE:RESults:FORMat:CURSor" on page 839
• ":SAVE:RESults:FORMat:MASK" on page 840
• ":SAVE:RESults:FORMat:SEARch" on page 842
• ":SAVE:RESults:FORMat:SEGMented" on page 843