1144 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:USBPd:TRIGger:HEADer:DMESsage
(see page 1610)
Command Syntax
:SBUS<n>:USBPd:TRIGger:HEADer:DMESsage <type>
<type> ::= {SRCap | REQuest | BIST | SNCap | BSTatus | ALERt | GCINfo
| VDEFined}
When the Header Content trigger mode is selected (with the
":SBUS<n>:USBPd:TRIGger HEADer" command) and the Data Message header type
is selected (with the ":SBUS<n>:USBPd:TRIGger:HEADer DMESsage" command),
the :SBUS<n>:USBPd:TRIGger:HEADer:DMESsage command selects the data
message type:
SRCap — Source_Capabilities
REQuest — Request
•BIST BIST
SNCap — Sink_Capabilities
BSTatus — Battery_Status
•ALERt Alert
GCINfo — Get_Country_Info
VDEFined — Vendor_Defined
Query Syntax
:SBUS<n>:USBPd:TRIGger:HEADer:DMESsage?
The :SBUS<n>:USBPd:TRIGger:HEADer:DMESsage? query returns the selected
data message type.
Return Format
<type><NL>
<type> ::= {SRC | REQ | BIST | SNC | BST | ALER | GCIN | VDEF}
See Also ":SBUS<n>:USBPd:SOURce" on page 1138
":SBUS<n>:USBPd:TRIGger" on page 1139
":SBUS<n>:USBPd:TRIGger:HEADer" on page 1140
":SBUS<n>:USBPd:TRIGger:HEADer:CMESsage" on page 1142
":SBUS<n>:USBPd:TRIGger:HEADer:EMESsage" on page 1145
":SBUS<n>:USBPd:TRIGger:HEADer:VALue" on page 1147
":SBUS<n>:USBPd:TRIGger:HEADer:QUALifier" on page 1148