1110 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:TYPE
(see page 1610)
Command Syntax
:SBUS<n>:UART:TRIGger:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
The :SBUS<n>:UART:TRIGger:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the :SBUS<n>:UART:TRIGger:DATA
and :SBUS<n>:UART:TRIGger:QUALifier commands are used to specify the data
value and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and alert bit
values) are only valid when a 9-bit width has been selected.
Query Syntax
:SBUS<n>:UART:TRIGger:TYPE?
The :SBUS<n>:UART:TRIGger:TYPE? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
TSTO | TDAT | TD1 | TD0 | TDX}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:MODE" on page 1308
• ":SBUS<n>:UART:TRIGger:DATA" on page 1107
• ":SBUS<n>:UART:TRIGger:QUALifier" on page 1109
• ":SBUS<n>:UART:WIDTh" on page 1111