1478 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:VOLTage:HIGH
(see page 1610)
Command Syntax
:WGEN<w>:VOLTage:HIGH <high>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<high> ::= high-level voltage in volts, in NR3 format
For all waveforms except DC, the :WGEN<w>:VOLTage:HIGH command specifies
the waveform's high-level voltage. Use the :WGEN<w>:VOLTage:LOW command
to specify the low-level voltage.
You can also specify the high-level and low-level voltages indirectly using the
:WGEN<w>:VOLTage and :WGEN<w>:VOLTage:OFFSet commands. For example, a
high-level voltage of 4 V and a low-level voltage of -1 V is the same as an
amplitude of 5 V and an offset of 1 V.
Query Syntax
:WGEN<w>:VOLTage:HIGH?
The :WGEN<w>:VOLTage:HIGH? query returns the currently specified waveform
high-level voltage.
Return Format
<high><NL>
<high> ::= high-level voltage in volts, in NR3 format
See Also "Introduction to :WGEN<w> Commands" on page 1435
":WGEN<w>:FUNCtion" on page 1446
":WGEN<w>:VOLTage:LOW" on page 1479
":WGEN<w>:VOLTage" on page 1477
":WGEN<w>:VOLTage:OFFSet" on page 1480
NOTE
When the amplitude is below 40 mV, the offset is limited to ±500 mV.