:CHANnel<n> Commands 11
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 363
:CHANnel<n>:RANGe
(see page 1610)
Command Syntax
:CHANnel<n>:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:RANGe command defines the full-scale vertical axis of the
selected channel. When using 1:1 probe attenuation, legal values for the range are
from 8 mV to 40 V.
If the probe attenuation is changed, the range value is multiplied by the probe
attenuation factor.
Query Syntax
:CHANnel<n>:RANGe?
The :CHANnel<n>:RANGe? query returns the current full-scale range setting for
the specified channel.
Return Format
<range_argument><NL>
<range_argument> ::= vertical full-scale range value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 345
":CHANnel<n>:SCALe" on page 364
":CHANnel<n>:PROBe" on page 354
Example Code
' CHANNEL_RANGE - Sets the full scale vertical range in volts. The
' range value is 8 times the volts per division.
myScope.WriteString ":CHANnel1:RANGe 8" ' Set the vertical range to
8 volts.
See complete example programs at: Chapter 44, “Programming Examples,”
starting on page 1619

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals