:WGEN<w> Commands 37
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1473
:WGEN<w>:TRACk:AMPLitude
(see page 1610)
Command Syntax
:WGEN<w>:TRACk:AMPLitude <on_off>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<on_off> ::= {{1 | ON} | {0 | OFF}
The :WGEN<w>:TRACk:AMPLitude ON command causes amplitude and offset
adjustments to this waveform generator output signal to be tracked by the other
waveform generator output.
Also, when tracking is enabled, adjustments for tracked settings in the other
waveform generator become unavailable (ghosted).
Query Syntax
:WGEN<w>:TRACk:AMPLitude?
The :WGEN<w>:TRACk:AMPLitude? query returns whether amplitude tracking is
on or off.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also • ":WGEN<w>:TRACk" on page 1472
• ":WGEN<w>:TRACk:FREQuency" on page 1475
• ":WGEN<w>:TRACk:PHASe" on page 1476
• ":WGEN<w>:TRACk:CSIGnal" on page 1474