716 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:CLResponse:DATA:GMARgin
(see page 1610)
Query Syntax
:POWer:CLResponse:DATA:GMARgin?
After the Control Loop Response (Bode) power analysis has been performed (see
:POWer:CLResponse:APPLy), the :POWer:CLResponse:DATA:GMARgin? query
returns the gain margin in dB.
Return Format
<gain_margin><NL>
<gain_margin> ::= gain margin in dB in NR3 format.
The query returns +9.9E+37 if the value cannot be calculated from the last sweep
(that is, if there are no crossover points).
See Also • ":POWer:CLResponse:APPLy" on page 714
• ":POWer:CLResponse:DATA" on page 715
• ":POWer:CLResponse:DATA:GMARgin:FREQuency" on page 717
• ":POWer:CLResponse:DATA:PMARgin" on page 718
• ":POWer:CLResponse:DATA:PMARgin:FREQuency" on page 719