:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1105
:SBUS<n>:UART:TRIGger:BASE
(see page 1610)
Command Syntax
:SBUS<n>:UART:TRIGger:BASE <base>
<base> ::= {ASCii | HEX}
The :SBUS<n>:UART:TRIGger:BASE command sets the front panel UART/RS232
trigger setup data selection option:
• ASCii — front panel data selection is from ASCII values.
• HEX — front panel data selection is from hexadecimal values.
The :SBUS<n>:UART:TRIGger:BASE setting does not affect the
:SBUS<n>:UART:TRIGger:DATA command which can always set data values using
ASCII or hexadecimal values.
Query Syntax
:SBUS<n>:UART:TRIGger:BASE?
The :SBUS<n>:UART:TRIGger:BASE? query returns the current UART base setting.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:MODE" on page 1308
• ":SBUS<n>:UART:TRIGger:DATA" on page 1107
NOTE
The :SBUS<n>:UART:TRIGger:BASE command is independent of the :SBUS<n>:UART:BASE
command which affects decode and Lister only.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals