1016 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:MANChester:TSIZe
(see page 1610)
Command Syntax
:SBUS<n>:MANChester:TSIZe <#bits>
<#bits> ::= from 0-32, in NR1 format
When the display format is WORD (see :SBUS<n>:MANChester:DISPlay), the
:SBUS<n>:MANChester:TSIZe command specifies the number of bits in the trailer
field of your Manchester protocol definition.
Query Syntax
:SBUS<n>:MANChester:TSIZe?
The :SBUS<n>:MANChester:TSIZe? query returns the number of trailer field bits
setting.
Return Format
<#bits><NL>
<#bits> ::= from 0-32, in NR1 format
See Also ":SBUS<n>:MANChester:DISPlay" on page 1004
":SBUS<n>:MANChester:DSIZe" on page 1005
":SBUS<n>:MANChester:HSIZe" on page 1006
":SBUS<n>:MANChester:SSIZe" on page 1010
":SBUS<n>:MANChester:WSIZe" on page 1017