:COMPliance Commands 12
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 371
:COMPliance:USB:SOURce:ADJacent
(see page 1610)
Command Syntax
:COMPliance:USB:SOURce:ADJacent <source>
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :COMPliance:USB:SOURce:ADJacent command specifies the analog input
channel that is probing the adjacent D+ or D- signal.
When the Device Full Speed Signal Quality test is selected, the specified channel
probes the adjacent D+ signal.
When the Device Low Speed Signal Quality test is selected, this specified channel
probes the adjacent D- signal.
Query Syntax
:COMPliance:USB:SOURce:ADJacent?
The :COMPliance:USB:SOURce:ADJacent? query returns the specified analog
input channel.
Return Format
<source><NL>
<source> ::= {CHAN<n>}
See Also • ":COMPliance:USB:SOURce:DPLus" on page 374
• ":COMPliance:USB:SOURce:DMINus" on page 373

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals