:SEARch Commands 32
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1201
:SEARch:SERial:FLEXray:MODE
(see page 1610)
Command Syntax
:SEARch:SERial:FLEXray:MODE <value>
<value> := {FRAMe | CYCLe | DATA | HERRor | FERRor | AERRor}
The :SEARch:SERial:FLEXray:MODE command selects the type of FlexRay
information to find in the Lister display:
FRAMe — searches for FlexRay frames with the specified frame ID.
CYCLe — searches for FlexRay frames with the specified cycle number and
frame ID.
DATA — searches for FlexRay frames with the specified data, cycle number, and
frame ID.
HERRor — searches for header CRC errors.
FERRor — searches for frame CRC errors.
AERRor — searches for all errors.
Query Syntax
:SEARch:SERial:FLEXray:MODE?
The :SEARch:SERial:FLEXray:MODE? query returns the currently selected mode.
Return Format
<value><NL>
<value> := {FRAM | CYCL | DATA | HERR | FERR | AERR}
See Also Chapter 32, “:SEARch Commands,” starting on page 1149
":SEARch:SERial:FLEXray:FRAMe" on page 1200
":SEARch:SERial:FLEXray:CYCLe" on page 1197
":SEARch:SERial:FLEXray:DATA" on page 1198

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals