676 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
26 :MTESt Commands
:MTESt:COUNt:FWAVeforms
(see page 1610)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and all
waveforms collected on the channel specified by the optional parameter or
collected on the currently specified source channel (:MTESt:SOURce) if there is no
parameter.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also • "Introduction to :MTESt Commands" on page 667
• ":MTESt:COUNt:WAVeforms" on page 679
• ":MTESt:COUNt:TIME" on page 678
• ":MTESt:COUNt:RESet" on page 677
• ":MTESt:SOURce" on page 697
Example Code • "Example Code" on page 667