1520 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
39 Obsolete and Discontinued Commands
:MEASure:TDELta
(see page 1610)
Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the Tstop
marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at the stop
marker (X2 cursor). No measurement is made when the :MEASure:TDELta? query
is received by the oscilloscope. The delta time value that is output is the current
value. This is the same value as the front-panel cursors delta X value.
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also "Introduction to :MARKer Commands" on page 529
"Introduction to :MEASure Commands" on page 568
":MARKer:X1Position" on page 533
":MARKer:X2Position" on page 536
":MARKer:XDELta" on page 538
":MEASure:TSTArt" on page 1524
":MEASure:TSTOp" on page 1525
NOTE
The :MEASure:TDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:XDELta command (see page 538) instead.