EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #430 background imageLoading...
Page #430 background image
430 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
17 :DVM Commands
:DVM:ENABle
(see page 1610)
Command Syntax
:DVM:ENABle <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
The :DVM:ENABle command turns the digital voltmeter (DVM) analysis feature on
or off.
Query Syntax
:DVM:ENABle?
The :DVM:ENABle? query returns a flag indicating whether the digital voltmeter
(DVM) analysis feature is on or off.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":DVM:SOURce" on page 432
• ":DVM:MODE" on page 431
• ":DVM:ARANge" on page 428

Table of Contents