:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1077
:SBUS<n>:SPI:CLOCk:TIMeout
(see page 1610)
Command Syntax
:SBUS<n>:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR3 format
The :SBUS<n>:SPI:CLOCk:TIMeout command sets the SPI signal clock timeout
resource in seconds from 100 ns to 10 s when the :SBUS<n>:SPI:FRAMing
command is set to TIMeout. The timer is used to frame a signal by a clock timeout.
Query Syntax
:SBUS<n>:SPI:CLOCk:TIMeout?
The :SBUS<n>:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also "Introduction to :TRIGger Commands" on page 1295