1458 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:MODulation:FSKey:RATE
(see page 1610)
Command Syntax
:WGEN<w>:MODulation:FSKey:RATE <rate>
<w> ::= 1 in NR1 format
<rate> ::= FSK modulation rate in Hz in NR3 format
The :WGEN<w>:MODulation:FSKey:RATE command specifies the rate at which the
output frequency "shifts".
The FSK rate specifies a digital square wave modulating signal.
Query Syntax
:WGEN<w>:MODulation:FSKey:RATE?
The :WGEN<w>:MODulation:FSKey:RATE? query returns the FSK rate setting.
Return Format
<rate><NL>
<rate> ::= FSK modulation rate in Hz in NR3 format
See Also • ":WGEN<w>:MODulation:AM:DEPTh" on page 1453
• ":WGEN<w>:MODulation:AM:FREQuency" on page 1454
• ":WGEN<w>:MODulation:FM:DEViation" on page 1455
• ":WGEN<w>:MODulation:FM:FREQuency" on page 1456
• ":WGEN<w>:MODulation:FSKey:FREQuency" on page 1457
• ":WGEN<w>:MODulation:FUNCtion" on page 1459
• ":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1460
• ":WGEN<w>:MODulation:STATe" on page 1462
• ":WGEN<w>:MODulation:TYPE" on page 1463