:MTESt Commands 26
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 683
:MTESt:LOCK
(see page 1610)
Command Syntax
:MTESt:LOCK <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:LOCK command enables or disables the mask lock feature:
ON — Locks a mask to the SOURce. As the vertical or horizontal scaling or
position of the SOURce changes, the mask is redrawn accordingly.
OFF — The mask is static and does not move.
Query Syntax
:MTESt:LOCK?
The :MTESt:LOCK? query returns the current mask lock setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 667
":MTESt:SOURce" on page 697