728 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:CLResponse:WGEN:LOAD
(see page 1610)
Command Syntax
:POWer:CLResponse:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :POWer:CLResponse:WGEN:LOAD command sets the waveform generator
expected output load impedance.
The output impedance of the Gen Out signal is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load. If the actual load impedance is
different than the selected value, the displayed amplitude and offset levels will be
incorrect.
Query Syntax
:POWer:CLResponse:WGEN:LOAD?
The :POWer:CLResponse:WGEN:LOAD? query returns the waveform generator
expected output load impedance setting.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also • ":POWer:CLResponse" on page 713
• ":POWer:CLResponse:APPLy" on page 714
• ":POWer:CLResponse:DATA" on page 715
• ":POWer:CLResponse:FREQuency:MODE" on page 720
• ":POWer:CLResponse:FREQuency:STARt" on page 722
• ":POWer:CLResponse:FREQuency:STOP" on page 723
• ":POWer:CLResponse:PPDecade" on page 724
• ":POWer:CLResponse:SOURce:INPut" on page 725
• ":POWer:CLResponse:SOURce:OUTPut" on page 726
• ":POWer:CLResponse:WGEN:VOLTage" on page 729
• ":POWer:CLResponse:WGEN:VOLTage:PROFile" on page 730