838 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
30 :SAVE Commands
:SAVE:RESults:[STARt]
(see page 1610)
Command Syntax
:SAVE:RESults:[STARt] [<file_spec>]
<file_name> ::= quoted ASCII string
The :SAVE:RESults:[STARt] command saves analysis results to a
comma-separated values (*.csv) file on a USB storage device.
Use the :SAVE:RESults:FORMat commands to specify the analysis types whose
results are saved to the file.
When multiple types of analysis results are selected, they are all saved to the same
file and separated by a blank line.
See Also • ":SAVE:RESults:FORMat:CURSor" on page 839
• ":SAVE:RESults:FORMat:MASK" on page 840
• ":SAVE:RESults:FORMat:MEASurement" on page 841
• ":SAVE:RESults:FORMat:SEARch" on page 842
• ":SAVE:RESults:FORMat:SEGMented" on page 843