1518 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
39 Obsolete and Discontinued Commands
:MEASure:LOWer
(see page 1610)
Command Syntax
:MEASure:LOWer <voltage>
The :MEASure:LOWer command sets the lower measurement threshold value. This
value and the UPPer value represent absolute values when the thresholds are
ABSolute and percentage when the thresholds are PERCent as defined by the
:MEASure:DEFine THResholds command.
Query Syntax
:MEASure:LOWer?
The :MEASure:LOWer? query returns the current lower threshold level.
Return Format
<voltage><NL>
<voltage> ::= the user-defined lower threshold in volts in NR3 format
See Also • "Introduction to :MEASure Commands" on page 568
• ":MEASure:THResholds" on page 1521
• ":MEASure:UPPer" on page 1527
The :MEASure:LOWer command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 577)
instead.