:TRIGger Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1367
:TRIGger:SHOLd:SOURce:CLOCk
(see page 1610)
Command Syntax
:TRIGger:SHOLd:SOURce:CLOCk <source>
<source> ::= {CHANnel<n> | DIGital<d>}
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger:SHOLd:SOURce:CLOCk command selects the input channel probing
the clock signal.
Query Syntax
:TRIGger:SHOLd:SOURce:CLOCk?
The :TRIGger:SHOLd:SOURce:CLOCk? query returns the currently set clock signal
source.
Return Format
<source><NL>
<source> ::= {CHAN<n> | DIG<d>}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:MODE" on page 1308
• ":TRIGger:SHOLd:SLOPe" on page 1366