:WAVeform Commands 36
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1421
:WAVeform:SOURce:SUBSource
(see page 1610)
Command Syntax
:WAVeform:SOURce:SUBSource <subsource>
<subsource> ::= {{SUB0 | RX | MOSI | FAST}
| {SUB1 | TX | MISO | SLOW}}
If the :WAVeform:SOURce is SBUS<n> (serial decode), more than one data set
may be available, and this command lets you choose from the available data sets.
When using UART serial decode, this option lets you get "TX" data. (TX is an alias
for SUB1.) The default, SUB0, specifies "RX" data. (RX is an alias for SUB0.)
When using SPI serial decode, this option lets you get "MISO" data. (MISO is an
alias for SUB1.) The default, SUB0, specifies "MOSI" data. (MOSI is an alias for
SUB0.)
When using SENT serial decode, this option lets you get "SLOW" data. (SLOW is
an alias for SUB1.) The default, SUB0, specifies "FAST" data. (FAST is an alias for
SUB0.)
If the :WAVeform:SOURce is not SBUS, or the :SBUS<n>:MODE is not UART, SPI,
or SENT, the only valid subsource is SUB0.
Query Syntax
:WAVeform:SOURce:SUBSource?
The :WAVeform:SOURce:SUBSource? query returns the current waveform
subsource setting.
Return Format
<subsource><NL>
<subsource> ::= {SUB0 | SUB1}
See Also • "Introduction to :WAVeform Commands" on page 1396
• ":WAVeform:SOURce" on page 1417